Manómetro para la presión pmax = 60 mbar

RITTER · Made in Germany
Manómetro para la presión pmax = 60 mbar · Ficha de datos

Table of Contents

Overview

Measuring Range 0 to 60 mbar
Resolution 2 mbar

Application

The Manometer can be used for measurement of the gas pressure while measuring the gas flow. Among other reasons, this is necessary if the measured and indicated actual volume of gas must be recalculated into the norm volume. The actual volume is the volume at the actual temperature and the actual pressure.

The norm volume of a gas is the volume at norm conditions which are (in Germany):

Norm Temperature = 273.15 Kelvin ( = 0 °C)
Norm Pressure = 1,013.25 mbar

The formula for converting the actual volume into norm volume is:

VN = Norm Volume in (ltr)
Vi = Indicated Volume in (ltr)
pN = Norm Pressure in (mbar–absolute)
pa = Actual Pressure in (mbar–absolute)
TN = Norm Temperature in (Kelvin)
Ti = Indicated Temperature in (Kelvin)

Note: The indicated gas pressure at the manometer is the differential pressure between the gas pressure at the gas inlet and the actual atmospheric air pressure. Thus, the actual gas pressure (pa) of the above formula equals the indicated gas pressure at the Manometer plus the actual atmospheric air pressure in (mbar).

Installation

Unpack the Manometer which is mounted into a T-piece. According to the rules for calibration and measurement with gas meters, the Manometer must be positioned at the gas inlet of the meter. The gas inlet nozzle is labelled accordingly.
Mount the Manometer onto the gas inlet nozzle by tightly screwing the union nut which is attached to the Thermometer. Thus, the Manometer is ready for use.