TG 0.5 Stainless steel

RITTER · Made in Germany
TG 0.5 Stainless steel · Data Sheet

Performance Data

Minimum flow Qmin 1 ltr/h
Standard flow Qstandard 50 ltr/h
Maximum flow Qmax 60 ltr/h
Measurement accuracy at standard flow 0.2 %
Measurement accuracy across measuring range 0.5 %
Measuring drum volume 0.5 ltr/U
Packing liquid quantity, approx 3.5 ltr
Maximum gas inlet pressure (1) 500 mbar
Minimum differential pressure (2) 0.4 mbar
Minimum dial division 0.002 ltr
Maximum indication value (3) 99,999,999 ltr
Connection gas in/outlet
Inside-Ø 10.6 mm
Outside-Ø 14 mm
(1) Standard version: 500 mbar – special version also available for 1 bar
(2) Differential pressure (= pressure loss) gas inlet gas outlet
(3) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
1 Stainless Steel PVC-grey 265 mm 310 mm 380 mm 8.3 kg 40 °C
2 Stainless Steel PE-el 265 mm 310 mm 380 mm 8.2 kg 60 °C
3 Stainless Steel PP-grey 265 mm 310 mm 380 mm 8.2 kg 80 °C
4 Stainless Steel PVDF 265 mm 310 mm 380 mm 8.5 kg 80 °C

Caution

Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion!

For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • 8-digit Totalizing Roller Counter
  • Acoplamento magnético entre o tambor de medição e o contador: Invólucro absolutamente estanque a líquidos
  • Bubble level for levelling
  • Level and Levelling Feet
  • Multi-Chamber Measuring Drum
  • Supports for Thermometer and Manometer
  • Viton sealing

V 3.2 / Rev. 2023-03-09 / Subject to alterations.