Performance Data
Minimum flow Qmin | 100 | ltr/h |
Standard flow Qstandard | 10,000 | ltr/h |
Maximum flow Qmax | 18,000 | ltr/h |
Measurement accuracy at standard flow | ± 0.2 | % |
Measurement accuracy across measuring range | ± 0.5 | % |
Measuring drum volume | 50 | ltr/U |
Packing liquid quantity, approx | 91 | ltr |
Maximum gas inlet pressure | 50 | mbar |
Minimum differential pressure (1) | 0.1 | mbar |
Minimum dial division | 0.5 | ltr |
Maximum indication value (2) | 999,999,990 | ltr |
Connection gas in/outlet | ||
Inside-Ø | 31.8 | mm |
Outside-Ø | 40 | mm |
(1) Differential pressure (= pressure loss) gas inlet gas outlet
(2) Standard Totalizing Roller Counter
Models (materials)
Caution
Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion!For chemical resistance properties please contact your RITTER distribution partner!
Standard Equipment
- 8-digit Totalizing Roller Counter
- Acoplamento magnético entre o tambor de medição e o contador: Invólucro absolutamente estanque a líquidos
- Bubble level for levelling
- Level and Levelling Feet
- Multi-Chamber Measuring Drum
- Supports for Thermometer and Manometer
- Viton sealing
Built-in Options
- LCD Display
- Pulse Generator
- Indicador de Nível de Líquido de Alta Precisão para Embalagem »HPLI« , para o nível de líquido de embalagem
V 2.1 / Rev. 2023-03-09 / Subject to alterations.
The most recent version of this data-sheet can be found at …
https://dlr.ritter.de/pt-br/tg-50-plastic/
Dr.-Ing. RITTER Apparatebau GmbH & Co. KG · Coloniastrasse 19-23 · D-44892 Bochum · Germany
For questions please contact mailbox@ritter.de or your local distribution partner (on our overview page)
https://dlr.ritter.de/pt-br/tg-50-plastic/
Dr.-Ing. RITTER Apparatebau GmbH & Co. KG · Coloniastrasse 19-23 · D-44892 Bochum · Germany
For questions please contact mailbox@ritter.de or your local distribution partner (on our overview page)