Oxygen Sensor

RITTER · Made in Germany MultiGas
Oxygen Sensor · Data Sheet

Table of Contents

Overview

  • Electrochemical sensor
  • Measuring ranges:
  • Standard version 0 – 25 % or 0 – 100 %
  • H2S resistant version 0.5 – 35 %
  • Measurement accuracy: ±2 % of span (full scale)
  • Resolution: < 0.5 % of span (full scale)
  • Response time (t90): 5 – 10 s; automotive version < 3.5 s
  • Lifetime: approx. 5 years
The oxygen sensor is a sensor module available as option in addition to a RITTER MultiGas NDIR or NDUV sensor. The measured oxygen concentration is displayed in the provided software. The sensor is built into the casing of the RITTER MultiGas sensors.

The following versions are available:

  • Standard version suitable for non-aggressive gases
  • H2S and similar acidic gases resistant version

Specifications

General features

VersionStandard versionH2S resistant version
Measurement range0 – 25 Vol.% O20 – 100 Vol.% O20.5 – 35 Vol.% O2
ApplicationBiogas, Automotive exhaust gas analyserIndustrial, fully CO2 resistantIndustrial, fully CO2 resistant,
shows high resistance to acid gases
Medium contact materialsABS, FKM, PPS, PTFE, stainless steelABS, PVC, PPS, PTFE, stainless steelABS, PVC, PPS, PTFE, stainless steel
Expected operating life1,000,000 Vol.% O2 h~ 1,200,000 Vol.% O2 h~ 1,200,000 Vol.% O2 h
Sensor lifetime4 years at ambient air,
depending on application
6 years at ambient air,
depending on application
6 years at ambient air
Dimensions (H x W x L)65.4 mm × 31.7 mm × 56.6 mm
Weight70 g
Tube connector4/6 mm tube

Measuring response *

VersionStandard versionH2S resistant version
Measurement range0 – 25 Vol.% O20 – 100 Vol.% O20.5 – 35 Vol.% O2
Resolution0.1 Vol.%0.1 Vol.%0.1 Vol.%
Response time (t90)< 3.5 s< 10 s< 5 s
Drift **< 1% per month< 1% per month< 3% per month
Linearity Error0 – 2 Vol.% O2: ± 0.1 abs.
2.1 – 100 Vol.% O2: ± 0.05 rel.
0 – 2 Vol.% O2: ± 0.1 abs.
2.1 – 35 Vol.% O2: ± 0.05 rel.
Repeatability ***± 1 Vol.% O2± 1 Vol.% O2
Influence of Humidity–0.03 % rel. O2 reading / % RH–0.03 % rel. O2 reading / % RH–0.03 % rel. O2 reading / % RH
InterferencesCO2: up to 20 Vol.%
CO: up to 2000 ppm
NOx: up to 5000 ppm
HC: up to 5000 ppm
N2O: up to 500 ppm
< 20 ppm O2 response to:
100 Vol.% CO
100 Vol.% CO2
100 Vol.% C3H8
3000 ppm NO in N2
1000 ppm C6H6 in N2
500 ppm SO2 in N2
< 100 ppm O2 response to:
3000 ppm C2H6O
3000 ppm C4H10S
< 200 ppm O2 response to:
3000 ppm C2H6S2
< 400 ppm O2 response to:
100 Vol.% H2
< 500 ppm O2 response to:
2000 ppm H2S in N2
< 20 ppm O2 response to:
100 Vol.% CO
100 Vol.% CO2
100 Vol.% C3H8
1000 ppm C6H6 in N2
2000 ppm H2S in N2
< 20000 ppm O2 response to:
3000 ppm NO in N2
1000 ppm H2 in N2
500 ppm SO2 in N2

(*) related to Pa = 1013 hPa, Ta = 25 °C, RH = 50%, flow = 2.5l/min
(**) averaged across 12 months
(***) @ 100 Vol.% O2 applied for 5 min

Climatic conditions

VersionStandard versionH2S resistant version
Measurement range0 – 25 Vol.% O20 – 100 Vol.% O20.5 – 35 Vol.% O2
Operating temperature0 – 40 °C
intermittent 40 – 50 °C
0 – 45 °C0 – 50 °C
Storage temperature-20 – 40 °C
5 – 25 °C recommended
40 – 50 °C max. 1 week
5 – 30 °C recommended
-20 – 50 °C maximum
Air pressure650 – 1250 hPa (mbar)700 – 1250 hPa (mbar)600 – 1250 hPa (mbar)
Ambient humidity0 – 95% rel. humidity (not condensing)

V 1.0 / Rev. 2023-03-21 / Subject to alterations.