MultiGas Ultraviolet Module NDUV H2S

Specifications H2S Sensor

The H2S sensor works according to the principle of non-dispersive UV absorption (NDUV). The measuring wavelength used is in the lower nanometer range, which enables inter-ference-free measurements with water vapour and hydrocarbons. This makes the sensor ideal for use in biogas measurements, as accompanying gases such as NH3, CO2, CH4 and H2O do not affect the accuracy of the measurements.

General features

Measurement technology:Innovative NDUV Sensor
(non-dispersive ultraviolet sensor) Dual beam
Detectable gases:
H2S SO2
Number of simultaneously detectable gases:max. 2
Measurement ranges:See below
Flow rate range:6 ~ 300 ltr/h – For higher flow rates the sensor can be operated in bypass
Max. gas inlet pressure:300 mbar
Pressure loss (without additional optional sensors):10 @ 100 / 35 @ 200 / 70 @ 300 (mbar @ ltr/h)
Temperature compensation:Yes
Data acquisition software:Yes
Lifetime of UV radiation source:> 8 000 h
Measurement cuvette:Stainless steel with silicone coating inside
Cuvette sealing:Viton O-ring
Casing:High-quality table-top casing, aluminium
Dimensions (W x H x L):464 x 189 x 305 mm
Weight: approx. 6.5+ kg
Gas connections:PVDF screw-type tube connection for tube Øi 4mm, Øo 6 mm

Measuring response

Linearity error:< ± 1% F.S.
Repeatability:± 0.5% F.S.
Long term stability zero N2:< ± 1% F.S. / 24h
Long term stability span:< ± 1% F.S. / month
Temperature influence of zero point:< 1% F.S. / 10K
Temperature influence of span:< 2% F.S. / 10K
Cross sensitivity:< 2% F.S.
Pressure influence:< 1.5% / 10hPa of reading
Warm-up time:1 min (initial), < 15 min for full specification
Response time (t90):1.5 – 15 sec
Sampling frequency by software:≤ 10 Hz
Detection limit:See below
Resolution:0.5 x detection limit

Electrical features

Power supply: 24 VDC incl. power plug 100 ~ 240 VAC / 24 VDC
Supply current (peak):<0.4 A
Average power consumption:< 7.5 W
Interface:USB (standard), RS232 (option) – incl. data transmission cable 1 m
Analogue voltage output (option): 0-2V / 0-5V / 0-10V

Climatic conditions

Operating temperature: +5 ~ +45 °C
Storage temperature: –20 ~ +60 °C
Operating pressure: 800 ~ 1200 hPa (mbar)
Ambient humidity:0 ~ 95% rel. humidity // Condensing inside of sensor must be prevented!